Sources Of Error With Use Of Goldmann Applanation Tonometers

Nilüfer Koçak*, Üzeyir Günenç**
* Araş.Gör.Dr.,Dokuz Eylül Üniversitesi Tıp Fakültesi Göz Hastalıkları AD,** Doç.Dr.,Dokuz Eylül Üniversitesi Tıp Fakültesi Göz Hastalıkları AD, İZMİR



There are many sources of error in the use of Goldmann- type applanation tonometers. Clinical conditions that can cause underestimation of intraocular pressure include valsalva maneuver, corneal epithelial edema, accommodation, repeated tonometry, hypofluorescence of the precorneal tear film, and conditions in which the tonometer is in excess contact with precorneal tear film. Measurements made over the surface of contact lens, increased stromal thickness, paracentral placement of tonometer tip can result in overestimation of intraocular pressure. In order to minimize error rate during the measurement procedure, one should employ extreme caution. The routine care and calibration of the device is also of utmost importance.

Keywords: Goldmann applanation tonometer, Intraocular pressure, errors in tonometric readings

Turkiye Klinikleri J Ophthalmol 1999, 8:70-72

E-mail To Friend This Page Printer Friendly Page
 
 
 ARAMA
  
 Ayrıntılı Arama
 GİRİŞ
 Kullanıcı Adı :
 
 Şifre  
   Beni Hatırla
 Şifremi Unuttum
 Kayıt Ol
 
 
 
 
 
 
 
 

Yasal uyarı: Bu sitede yayınlanan resim, yazı ve diğer uygulamaların her hakkı Ortadoğu Reklam Ve Yayıncılık A.Ş. 'ye aittir. Kaynak gösterilmeden kullanılamaz. Bu site hekimleri sağlık alanında bilgilendirmeye yönelik hazırlanmıştır. Sitede yer alan bilgiler tanı ve tedavi amaçlı kullanıldığında sorumluluk tamamen kullanıcıya aittir. Siteye girmekle bu şartları okumuş, anlamış ve kabul etmiş sayılırsınız.